53

Press the “Slope” option button to select “

” , “

” or “

”.

4. Set up Trigger mode

Press the “Trigger mode” option button to select “Auto”,

“Normal”, “Single”.

Auto: The waveform refreshes at a high speed whether

the trigger condition is satisfied or not.

Normal: The waveform refreshes when the trigger

condition is satisfied and waits for next trigger event

occurring when the trigger condition is not satisfied.

Single: The oscilloscope acquires a waveform when the

trigger condition is satisfied and then stops.

5. Set up Trigger coupling

Press the “Set up” button to enter the “Trigger Setup

Menu”.

Press the “Coupling” option button to select “DC”,

“AC”, “HF Reject” or “LF Reject”.

Pulse Trigger

Use Pulse Width triggering to trigger on aberrant pulses.

Table 16 - Pulse trigger Function Menu 1

Option

Setting

Description

Type

Pulse

Select the pulse to trigger

the pulse matching the

trigger condition.

Source

CH1

CH2

*CH3

*CH4

EXT

EXT/5

Select input signal

source.