B&K Precision 2552-2559 - Manual User Manual

Page 140

Advertising
background image

140

6. Press the “Operation” option button and select “-”.

7. Press the “CH1-CH2” option button to display a new waveform

that is the difference between the displayed waveforms.

8. You can adjust the vertical scale and position of the Math

waveform. To do so, follow these steps:

a. Remove the channel 1 and channel 2 waveforms from the

display.

b. Turn the CH 1 and CH 2 VOLTS/DIV and VERTICAL

POSITION knobs to adjust the vertical scale and position.

NOTE. First be sure to compensate both probes. Differences in probe
compensation appear as errors in the differential signal.