X-y format – B&K Precision 2552-2559 - Manual User Manual

Page 79

Advertising
background image

79

Setup Grid

Press the “Grid” option button to select “

”, “

”or“

”to set the screen whether to

display grid or not.

Setup Menu Display

Press the “Menu Display” option button to select “2

sec”, “5 sec”, “10 sec”, “20 sec”, or “Infinite” to set

menu display time on screen.

Set Skin

Press the “skin” option button or turn the “Universal”

knob to select “Classical”, “Modern”, “Traditional”, or

“Succinct”.

X-Y Format

Use the XY format to analyze phase differences, such as those represented

by Lissajous patterns. The format plots the voltage on channel 1 against the

voltage on channel 2, where channel 1 is the horizontal axis and channel 2

is the vertical axis. The oscilloscope uses the untriggered Sample

acquisition mode and displays data as dots.

Operation steps:

Channel 1 (or channel 3) “Volts/div” and vertical

“POSITION” set up the horizontal scale and position.

Channel 2 (or channel 4) “Volts/div” and vertical

“POSITION” set up the horizontal scale and position.

Turn the “Time/div” knob to adjust the sampling rate

The following functions are forbidden at XY display

form:

● Benchmark waveform and wave mathematics

Cursor

Note: The oscilloscope can capture a waveform in normal YT mode

at any sampling rate. You can view the same waveform in XY mode.

To do so, stop the acquisition and change the display format to XY.