B&K Precision 2552-2559 - Manual User Manual
Page 145

145
factors
2552/2553 2554/2555
2556/2557
2558/2559
Vertical System
Vertical
Sensitivity
2 mV/div – 10 V/div (1-2-5 order)
Channel
voltage offset
range
2 mV–100 mV: ±800 mV
102 mV – 5 V: ±40 V
Vertical
Resolution
8 bit
Channels
2 (2552)
4 (2553)
2 (2554)
4 (2555)
2 (2556)
4 (2557)
2 (2558)
4 (2559)
Analog
Bandwidth (-
3dB)
70 MHz
100 MHz
200 MHz
300 MHz
Single-shot
Bandwidth
70 MHz
100 MHz
200 MHz
300 MHz
Lower
frequency
limit (AC -
3dB)
≤ 10 Hz (at BNC input)
DC Gain
Accuracy
< ± 3.0%: 5 mV/div to 5 V/div in fixed gain ranges
< ± 4.0%: 2 mV/div in variable gain ranges
DC
Measurement
Accuracy:
All Gain
settings
≤100mV/div
± [3%*(|reading|+|offset|)+1% *|offset| +0.2 div+2 mV]
DC
Measurement
Accuracy:
All Gain
settings
>100mV/div
± [3%*(|reading|+|offset|)+ 1%* offset| +0.2 div+100
mV]
Rise time
< 5 ns
< 3.5 ns
< 1.8 ns
< 1.2 ns