Chapter 6 Specifications

All specifications apply to measurements with a 10X probe. To

verify that the oscilloscope meets specifications, the oscilloscope

must first meet the following conditions:

o The oscilloscope must have been operating continuously

for thirty minutes within the specified operating

temperature.

o You must perform the Do Self Cal operation, accessible

through the Utility menu, if the operating temperature

changes by more than 5° C.

o The oscilloscope must be within the factory calibration

interval.

All specifications are guaranteed unless noted “typical.”

Specifications

2552/2553 2554/2555

2556/2557

2558/2559

Inputs

Input

Coupling

AC, DC, GND

Input

Impedance

1 MΩ ± 2% || 13 pf ± 3 pf 1 MΩ ± 2% || 13 pf ± 3 pf

50 Ω ± 2%

Maximum

input

voltage

400 V (DC+AC PK-PK, 1 MΩ input impedance, X10),

CAT I

Ch to Ch

Isolation

(Both

channels in

same V/div

setting)

>100:1 at

35 MHz

>100:1 at

50 MHz

>100:1 at

100 MHz

>100:1 at 150

MHz

Probe

attenuator

1X, 10X

Probe

attenuations

selectable

1X, 5X, 10X, 50X, 100X, 500X, 1000X