Updating firmware – B&K Precision 2552-2559 - Manual User Manual

Page 116

Advertising
background image

116

Figure 71 - LED Test Screen

Updating Firmware

Using USB flash drive to update firmware

The firmware of the oscilloscope can be updated directly via a USB Flash

drive. This process takes about two minutes. To update, follow the steps

below:

1. Download and save the firmware file onto a USB flash drive.

2. Insert the USB Flash Drive to the USB Host interface on the front

panel of the oscilloscope.

3. Press the “UTILITY” button to enter the “Utility menu”.

4. Press the “Next Page” option button to enter the third page of the

“Utility menu”.

5. Press the “Update Firmware” option button. Read the on-screen

notes carefully before proceeding.

6. Press the “SINGLE” button to enter into the directory menu of the

USB flash drive.

7. Use the universal knob to select the firmware file on the USB flash

drive, and select “Confirm” to start the update.

When completed, turn off the oscilloscope and turn on again. The software

will then be updated. The oscilloscope will require self calibration after the

update. Go into the utility menu’s second page and select “Do Self Cal” to

run self calibration.