signal to the EXT TRIG BNC and set the trigger source to Ext or

Ext/5.

“FORCE Button: Use the FORCE button to complete the current

waveform acquisition whether the oscilloscope detects a trigger or

not. This is useful for SINGLE acquisitions and Normal trigger

mode.

Signal Source

You can use the Trigger Source options to select the signal that the

oscilloscope uses as a trigger. The source can be any signal connected to a

channel BNC, to the EXT TRIG BUS, or the AC power line (available only

with Edge Trigger).

Trigger Type

The scopes have five trigger types: Edge, Video, Pulse, Slope, and

Alternate.

Edge Trigger

Use Edge triggering to trigger on the edge of the oscilloscope input signal

at the trigger threshold.

Table 14 - Edge Trigger Function Menu

Option

Setting

Description

Type

Edge

With Edge highlighted, the rising or falling edge

of the input signal is used for the trigger.

Source

CH1

CH2

*CH3

*CH4

Triggers on a channel whether or not the

waveform is displayed.