Default setup, Efault, Etup – B&K Precision 2552-2559 - Manual User Manual

Page 26

Advertising
background image

26

Note: The AUTO button can be disabled. Please see Appendix C

Disabling Auto Functionfor details.

Default Setup

The oscilloscope is set up for normal operation when it is shipped from the

factory. This is the default setup. To recall this setup, press the

DEFAULT SETUP button. For the default options, buttons, and controls,

when you press the DEFAULT SETUP button, refer to “Appendix A

Default Setup”.

The DEFAULT SETUP button does not reset the following settings:

Language option

Saved reference waveform files

Saved setup files

Display contrast

Calibration data

VOLTS/DIV

adjustability

Coarse

Signal inverted

Off

Horizontal position

Center

TIME/DIV

Adjusted

Trigger type

Edge

Trigger source

Auto detect the channel which has the input

signal

Trigger slope

Rising

Trigger mode

Auto

Trigger coupling

DC

Trigger holdoff

Minimum

Trigger level

Set to 50%