Application of x-y function, Pplication of, Unction – B&K Precision 2552-2559 - Manual User Manual

Page 138

Advertising
background image

138

3. Press the “Sync” option button and select “Line Num” and turn the

universal knob to set a specific line number.

4. Press the “Standard” option button and select “NTSC”.

5. Turn the “Time/Div” knob to see a complete video line across the

screen.

6. Turn the “Volts/Div” knob to ensure that the entire video signal is

visible on the screen.

Application of X-Y Function

Viewing Impedance Changes in a Network

Connect the oscilloscope to monitor the input and output of the circuit

To view the input and output of the circuit in an XY display, follow these

steps:

1. Press the “CH 1” MENU button and set the Probe option

attenuation to 10X.

2. Press the “CH 2” MENU button and set the Probe option

attenuation to 10X.

3. Set the switch of the probe to 10X.

4. Connect the channel 1 probe to the input of the network, and

connect the channel 2 probe to the output.

5. Press the “AUTO” button.

6. Turn the “Volts/div” knobs to display approximately the same

amplitude signals on each channel.

7. Press the “DISPLAY” button.

8. Press the Format option button and select “XY”.

9. The oscilloscope displays a Lissajous pattern representing the

input and output characteristics of the circuit.

10. Turn the Volts/div and VERTICAL POSITION knobs to optimize

the display.

11. Press the “Persist” option button and select “Infinite”.

12. Press the “Intensity” or “Brightness” option buttons and turn the

“Universal” knob to adjust the contrast of the screen.

Apply the Ellipse method to observe the phase difference between the two

channels