Chapter 5 message prompts and troubleshooting, Message prompts, Essage – B&K Precision 2552-2559 - Manual User Manual

Page 141: Rompts

Advertising
background image

141

Chapter 5 Message Prompts

and Troubleshooting

Message Prompts

Trig level at limit! : Prompts you that the trigger Level is at a

limit when you turn the Trig level knob.

Horizon position at limit! : Prompts you that the horizontal

position is at a limit when you turn the horizontal position knob.

Volts/Div at limit! : Prompts you that the vertical voltage has

already reached the Min 2mV/div or the Max 5V/div.

Volts position at limit! : The system would display this

information when the vertical position is at a limit.

Time/Div at limit! : Prompts that the Time/Div is at full range

while turning the horizontal scale knob.

Holdoff time at limit! : Prompts you that the holdoff time setting

has reached max or min value.

Function isn’t useable! : Under several special modes, some

functions could not be running.

No signal! : The system would display this information when the

signal could not match the auto set condition while using the auto

set function.

Adjust at limit! : Prompts that the pulse width setting has reached

min 20.0 ns or max 10.0 s.

Location Empty! : If you have no stored waveforms or setups at

specified location, the screen will display this information when

you press the “Recall” button on this location.

USB Flash Drive Plug In! : This information will appear when

you insert the USB Flash Drive to the USB Host port.

USB Flash Drive Pull Out! : This information will appear when

you pull out the USB Flash Drive from the USB Host port.

Store Data Success! : Saving setup data, waveform data or picture

data to the internal memory of the oscilloscope or USB flash was

successful.

Advertising