Measuring Ring Amplitude

To measure the amplitude, follow these steps:

1. Press the “CURSORS” button to see the Cursor Menu.

2. Press the “Mode” button to select “Manual”.

3. Press the “Type” option button and select “Voltage”.

4. Press the “Source” option button and select “CH1”.

5. Press the “CurA” option button and turn the universal knob to

place the cursor A on the highest peak of the ring.

6. Press the “CurB” option button and turn the universal knob to

place the cursor B on the lowest peak of the ring.

You can see the following measurements on the top of the left screen:

● The delta voltage (peak-to-peak voltage of the ringing)

● The voltage at Cursor A.

The voltage at Cursor B.

Figure 84 - Cursor Menu Screen 2

Capturing a single event

To capture a single event, you need to gather some knowledge of the signal

in order to correctly set up the trigger level and slope. If you are not sure

about the instance of the signal, you can observe it through auto or normal

trigger mode to ensure the trigger level and slope will capture the signal.