“GND” Coupling: Use GND coupling to display a zero-volt

waveform. Internally, the channel input is connected to a zero-volt

reference level.

Fine Resolution: The vertical scale readout displays the actual

Volts/Div setting while in the fine resolution setting. Changing the

setting to coarse does not change the vertical scale until the

VOLTS/DIV control is adjusted.

Option

Setting

Description

Filter

On

Off

Turn on the digital filter.

Turn off the digital filter.

Type

Setup as LPF (Low Pass Filter).

Setup as HPF (High Pass Filter).

Setup as BPF (Band Pass Filter).

Setup as BRF (Band Reject Filter).

Upp_limit

Turn the “Universal” knob to set upper limit.

Low_limit

Turn the “Universal” knob to set lower limit.

Return

Return to digital filter main menu.

Note: ● The oscilloscope’s vertical response rolls off slowly above its

specified bandwidth. Therefore, the FFT spectrum can show valid
frequency information higher than the oscilloscope’s bandwidth.
However, the magnitude information near or above the bandwidth
will not be accurate.
● If the channel is set to DC coupling, you can quickly measure the
DC component of the signal by simply noting its distance from the
ground symbol.
● If the channel is set to AC coupling, the DC component of the
signal is blocked allowing you to use greater sensitivity to display
the AC component of the signal.