Measure system, Scale measurement, Easure – B&K Precision 2552-2559 - Manual User Manual

Page 80: Ystem

Advertising
background image

80

● Auto (resets display format to YT)

Trigger Control

Horizontal Position Knob

Vector Display Type

Scan Display

Measure System

The oscilloscope displays the voltage in relation to time and tests the

waveform displayed. Different measurement techniques such as scale,

cursor, and auto measure modes are used.

Scale Measurement

This method allows you to make a quick visual estimate. For example, you

might look at waveform amplitude and determine that it is a little more than

100 mV. You can take simple measurements by counting the major and

minor graticule divisions involved and multiplying by the scale factor. For

example, if you counted five major vertical graticule divisions between

minimum and maximum values of a waveform and knew you had a scale

factor of 100 mV/div, then you could easily calculate your peak-to-peak

voltage as follows:

5 divisions x 100 mV/div = 500 mV

Note: When the display type is set to Vector, the oscilloscope

connects the sample points by using digital interpolation.

Interpolation mode contains linear interpolation and sin(x)/x

interpolation. Sin(x)/x interpolation is suitable for reconstructing

curved or irregular wave shapes, especially when the sample rate is 3

to 5 times the system bandwidth or at faster time base settings.