Probe Attenuation

Probes are available with various attenuation factors which affect the

vertical scale of the signal. The Probe Check function verifies that the

Probe attenuation option matches the attenuation of the probe.

You can push a vertical menu button (such as the CH 1 MENU button), and

select the Probe option that matches the attenuation factor of your probe.

Be sure that the attenuation switch on the probe matches the Probe option

in the oscilloscope. Switch settings are 1X and 10X.

Note: To avoid electric shock when using the probe, keep fingers behind

the guard on the probe body.

To avoid electric shock while using the probe, do not touch metallic

portions of the probe head while it is connected to a voltage source.
Connect the probe to the oscilloscope and connect the ground
terminal to ground before you take any measurements.

Note. The default setting for the Probe option is 1X.

Note. When the attenuation switch is set to 1X, the probe limits the

bandwidth of the oscilloscope to 10 MHz (according to Probe spec). To use

the full bandwidth of the oscilloscope, be sure to set the switch to 10X.