B&K Precision 2552-2559 - Manual User Manual

Page 143

Advertising
background image

143

“edge” trigger mode. The video signal should use the

“Video” Trigger mode. The signal would display steadily

only when using matching trigger mode.

3. Attempt to change the “coupling” into “HF Reject” or

“LF Reject” display, so that the High/low frequency

noise will be filtered.

5. When pressing “RUN/STOP” button, nothing displays.

Check whether or not the trigger mode on the trigger

interface is in “normal” or “single”, and then check the

trigger level to see if it is over the waveform range. If yes,

please move the trigger level to the middle position or set

the trigger mode to the “Auto” position. You can also press

the “Auto” button to set it up automatically.

6. After the Acquisition is set to Averages or Display Persistence

time is set too long, the waveform refreshes slowly.

This is a normal condition for these settings.

7. The signal is displayed as a ladder-like waveform.

1. This phenomenon is normal. The time base may be too

slow. Turn the horizontal SCALE knob to increase

horizontal resolution to improve the display.

2. Maybe the display Type is set to “Vectors”. You could

set it to Dots mode to improve the display.