GainCheck

Device self-test which runs automatically in

the background at fixed intervals. The

memory and signal transfer are checked.

You can also start the GainCheck manually.

Then, also a display test is performed and

the software version displayed.

Electrode zero

point

See Offset

Electrode slope

Is indicated in % of the theoretical slope

(59.2 mV/pH at 25 °C). The electrode slope is

different for each electrode and changes

with age and wear.

Offset

The voltage which a pH electrode provides

at a pH of 7. The offset is different for each

electrode and changes with age and wear.

One-point

calibration

Calibration with which only the offset (zero

point) is taken into account. The previous

slope value is maintained. Only one buffer

solution is required for a one-point

calibration.