Hach-Lange SI794 P User Manual User Manual
Page 77
77
Display
Action
Remark
Measuring mode:
From the blinking CAL mode
indicator you see that sample
calibration has not been
terminated.
Product calibration step 2:
When the sample value has
been determined, open the
product calibration once more
(cal, code 1105).
Display (approx.
3 sec)
Enter lab value.
The new offset is calculated.
Until the sample
value is deter-
mined and can be
entered, the
transmitter is in
measuring mode.
Display of slope and new offset
(related to 25°C).
Press enter to end calibration.
The measured value is shown
in the main display alternately
with “Hold”. Sensoface is active,
“enter” blinks.
Press enter to exit calibration.
After end of
calibration, the
outputs remain in
Hold mode for
approx. 20 sec.
New calibration:
Press cal.