Hach-Lange SI794 P User Manual User Manual

Page 77

Advertising
background image

77

Display

Action

Remark

Measuring mode:

From the blinking CAL mode

indicator you see that sample

calibration has not been

terminated.

Product calibration step 2:

When the sample value has

been determined, open the

product calibration once more

(cal, code 1105).

Display (approx.

3 sec)

Enter lab value.

The new offset is calculated.

Until the sample

value is deter-

mined and can be

entered, the

transmitter is in

measuring mode.

Display of slope and new offset

(related to 25°C).

Press enter to end calibration.

The measured value is shown

in the main display alternately

with “Hold”. Sensoface is active,

“enter” blinks.

Press enter to exit calibration.

After end of

calibration, the

outputs remain in

Hold mode for

approx. 20 sec.

New calibration:

Press cal.