Calibration error messages – Hach-Lange SI794 P User Manual User Manual

Page 89

Advertising
background image

89

Calibration error messages

Offset out of range (±60 mV)

• Electrode worn out

• Buffer solutions unusable or contaminated

Buffer does not belong to configured buffer set

Temperature probe not immersed in buffer

solution (for automatic temperature

compensation)

• Wrong buffer temperature set

(for manual temperature specification)

• Nominal electrode zero point ≠ pH 7

Icon blinks:

Problem

Possible causes

Problems during recognition

of the buffer solution

• Same or similar buffer solution was used for both

calibration steps

• Buffer solution used does not belong to buffer set

currently configured in the transmitter

Electrode slope out of range (80...103 %)

• Electrode worn out

• Buffer solutions unusable or contaminated

Buffer does not belong to configured buffer set

Temperature probe not immersed in buffer

solution (for automatic temperature

compensation)

Wrong buffer temperature set (for manual

temperature specification)

• Electrode used has different nominal slope