Stability check:

The measured mV value is

displayed.

To abort stability

check: Press cal.

(accuracy reduced)

Calibration with the first buffer

is terminated. Remove the

electrode and temp probe

from the first buffer solution

and rinse them thoroughly.

• One-point calibration:

Press cal to end the calibration.

Slope [%] and offset [mV] of

the electrode are displayed.

Press enter to proceed.

For one-point

calibration only:

• Two-point calibration:

Immerse electrode and

temperature probe in the

second buffer solution. Enter

pH value of second buffer

solution. Press enter to start.

The calibration

process runs again

as for the first buf-

fer.

Retract electrode and temp

probe out of second buffer,

rinse off, re-install.

Repeat calibration: cal
End calibration:

enter

Slope and offset of

electrode (related to

25 °C) are displayed.

pH value and “Hold” are

displayed alternately.

Sensoface is active, “enter”

blinks. Exit using enter key.

Hold is deactivated after 20 sec.