Temperature input *

Pt100/Pt1000/NTC 30 kΩ/NTC 8.55 kΩ/Balco 3kΩ

2-wire connection, adjustable

Measuring range

Pt 100/Pt 1000

-20.0 ... +200.0 °C

(-4 ... +392 °F)

NTC 30 kohms

-20.0 ... +150.0 °C

(-4 ... +302 °F)

NTC 8.55 kohms

-10.0 ... +130.0 °C

(-4 ... +266 °F)

Balco 3 kohms

0.0 ... +100.0 °C

(+32 ...+212 °F)

Adjustment range

10 K

Resolution

0.1 °C / 1 °F

Measurement error

1,2,3

<0.5 K (<1 K for Pt100; <1 K for NTC >100°C)

Temp compensation

Linear –19.99 ... +19.99 %/K

of process medium

(reference temp 25 °C)

HOLD input

Galvanically separated (OPTO coupler)

Function

Switches transmitter to HOLD mode

Switching voltage

0 ... 2 V (AC/DC)

Inactive

10 ... 30 V (AC/DC) Active

CONTROL input

Galvanically separated (OPTO coupler)

Function

Control input for automatic cleaning/calibration

system

Switching voltage

0 ... 2 V (AC/DC)

Inactive

10 ... 30 V (AC/DC) Active

Output 1

0/4 to 20 mA, max. 10 V, floating

(galvanically connected to output 2)

Measured variable *

pH or mV value

Overrange*

22 mA in the case of error messages

Output averaging filter *

Low-pass, filter time interval 0 ... 120 s

Measurement error

1

< 0.3% current value + 0.05 mA

Start/end of scale

Configurable within the measuring range for

pH or mV

Span allowed

pH 2.00 ... 18.00 / 200 ... 3000 mV