Menu code choices default – Hach-Lange SI794 P User Manual User Manual
Page 31

31
Menu
Code
Choices Default
ALrt
Alarm settings
AL.
Select Sensocheck
ON / OFF
OFF
Enter alarm delay
xxxx sec
0010 sec
LED in HOLD mode
ON / OFF
OFF
rLAY
Relay 1/2: Limit values, controller
rL.
Select limit function / controller
LiMIT / CtROL
LiMIT
Select contact function
Lo / Hi
Lo
L1.
Select contact response
N/O / N/C
N/O
Enter setpoint
xxxx
00.00 pH
Enter hysteresis
xxxx
00.50 pH
Enter delay
xxxx SEC
0010 SEC
Select contact function
Lo / Hi
Hi
L2.
Enter controller setpoint
xxxx
07.00 pH
Ct.
Enter neutral zone
xxxx
01.00 pH
(P) Controller gain K
P
xxxx %
0100 %
(I) Reset time T
R
xxxx SEC
0000 SEC
(D) Rate time T
D
xxxx SEC
0000 SEC
Pulse length/frequency controller
PLC / PFC
PLC
PLC: Pulse length
xxxx SEC
0010 SEC
PFC: Pulse frequency
xxxx /min
0060 /min
Select HOLD response
Y Last / Y Off
Y Last
Cln.
Cleaning contact
Cn.
Select cleaning contact or signaling
of parameter set 2
rinse / Pset2
rinse
rinse
Cleaning interval
xxx.x h
000.0 h
Cleaning duration
xxxx SEC
0060 SEC
Select contact response
N/O / N/C
N/O
Select contact response
N/O / N/C
N/O
Enter setpoint
xxxx
14.00 pH
Enter hysteresis
xxxx
00.50 pH
Enter delay
xxxx SEC
0010 sec