Diagnostics functions – Hach-Lange SI794 P User Manual User Manual

Page 81

Advertising
background image

81

Display of output currents

Press enter while in measuring mode.

The current at output 1 is shown in the main display,

the current at output 2 in the secondary display.

After 5 sec the transmitter returns to measuring mode.

Diagnostics functions

Display

Remark

Display of calibration data (Cal Info)

Press cal while in measuring mode and enter code:

0000. The slope is shown in the main display,

the offset in the secondary display.

After 20 sec the transmitter returns to measuring

mode (immediate return at pressing enter).

Display uncorrected electrode potential

(Sensor monitor)

Press conf while in measuring mode and enter code:

2222. The (uncompensated) electrode potential is

shown in the main display, the measuring tempera-

ture in the secondary display. Press enter to return to

measurement.

Display of last error message

(Error info)

Press conf while in measuring mode and enter code:

0000. The last error message is displayed for approx.

20 sec.

After that the message will be deleted (immediate

return to measurement at pressing enter).