Diagnostics functions – Hach-Lange SI794 P User Manual User Manual
Page 81

81
Display of output currents
Press enter while in measuring mode.
The current at output 1 is shown in the main display,
the current at output 2 in the secondary display.
After 5 sec the transmitter returns to measuring mode.
Diagnostics functions
Display
Remark
Display of calibration data (Cal Info)
Press cal while in measuring mode and enter code:
0000. The slope is shown in the main display,
the offset in the secondary display.
After 20 sec the transmitter returns to measuring
mode (immediate return at pressing enter).
Display uncorrected electrode potential
(Sensor monitor)
Press conf while in measuring mode and enter code:
2222. The (uncompensated) electrode potential is
shown in the main display, the measuring tempera-
ture in the secondary display. Press enter to return to
measurement.
Display of last error message
(Error info)
Press conf while in measuring mode and enter code:
0000. The last error message is displayed for approx.
20 sec.
After that the message will be deleted (immediate
return to measurement at pressing enter).