69

Display

Action

Remark

At the end of the adjustment

procedure the zero offset [mV]

of the electrode is displayed

(based on 25 °C).

Press enter to proceed.

This is not the final

calibration value of

the electrode! Zero

point and slope must

be determined using

a complete 2-point

calibration (cal 1100)

(see following pages).

Confirmation prompt.

Display of pH value (alternately

with Hold) and temperature,

“enter” blinks, Sensoface is

active.

Place electrode in process.

Press enter to end zero

calibration.

After end of calibra-

tion, the outputs

remain in Hold

mode for approx.

20 sec.

Note for zero adjustment

After having adjusted the nominal zero point, be sure to calibrate

the electrode following one of the procedures as described on

the next pages:

• Automatic calibration with Calimatic

• Manual calibration

• Data entry of premeasured electrodes