Display

Action

Remark

Enter the pH value of your

buffer solution for the proper

temperature. While the “hour-

glass” icon is blinking, the elec-

trode and temperature probe

remain in the buffer solution.

Manual Calibration

Automatic or manual temperature detection

Press cal, enter code: 1100

Select position using ,

edit number using ,

press enter to proceed.

Remove the electrode and

temperature probe, clean them

and immerse them in the first

buffer solution (in any order).

When “Manual temp detection”

has been configured, enter the

value in the secondary display

using the arrow keys.

Press enter to start.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

The response time

of the electrode

and temperature

probe is consider-

ably reduced when

the electrode is

first moved about

in the buffer solu-

tion and then held

still.

Transmitter in Hold

mode, measured

value frozen.

Sensoface inactive.

For calibration with manual buffer specification, you must enter the pH

value of the buffer solution used in the transmitter for the proper tempera-

ture. This presetting enables calibration with any desired buffer solution.

The MAN calibration mode and the type of temperature detection are

selected in the configuration mode.