Specifications – Hach-Lange SI794 P User Manual User Manual

Page 96

Advertising
background image

96

pH/mV input

Input for pH or ORP electrodes or ISFET sensors

Measuring range

-1500 ... +1500 mV

Display range

pH value

-2.00 ... 16.00

ORP

-1999 ... +1999 mV

Glass electrode input

1

Input resistance

> 0.5 x 1012 ohms

Input current

< 2 x 10-12 A

Reference electrode input

1

Input resistance

> 1 x 1010 ohms

Input current

< 1 x 10-10 A

Meas. error

1,2,3

pH value

< 0.02

TC: 0.002 pH/K (display)

mV value

< 1 mV

TC: 0.1 mV/K

pH sensor standardization *

pH calibration

Operating modes

BUF

Calibration with Calimatic automatic buffer

recognition

MAN

Manual calibration with input of individual

buffer values

DAT

Data entry of pre-measured electrodes

Zero adjustment

±200 mV

Max. calibration range

Offset: ±60 mV

Slope: 80 ... 103 % (47.5 ... 61 mV/pH)

ORP sensor standardization * ORP calibration

Max. calibration range

-700 ... +700 ∆mV

Cal timer

0000 ... 9999 h

Sensocheck

Automatic monitoring of glass and reference

electrode (can be disabled)

Sensoface

Provides information on the sensor condition,

evaluation of zero/slope, response time,

calibration interval, Sensocheck

Specifications