Display

Action

Remark

Select calibration (Press cal).

Enter passcode: 1100

Select position using ,

enter number using ,

press enter to proceed.

Remove the electrode and

temperature probe, clean them

and immerse them in the redox

buffer.

If an invalid code

is entered, the

transmitter returns

to measuring mode.

Display (approx.

3 sec)

The transmitter is

in the Hold mode.

Enter setpoint value for redox

buffer (secondary display:

electrode potential displayed

for approx. 6 sec)

Select position using ,

enter number using ,

press enter to proceed.

After approx. 6 sec

the secondary

display shows

the measured

temperature.

Display of electrode data

(delta value)

Press enter to proceed.

Rinse electrode and tempera-

ture probe and reinstall them.

“Zero” and “enter”

icons are blinking,

Sensoface is active.

The measured ORP value [mV]

is shown in the main display

alternately with “Hold”, “enter”

blinks, Sensoface is active.

Press enter to exit calibration.

After end of cali-

bration, the out-

puts remain in Hold

mode for approx.

20 sec.