Hach-Lange SI794 P User Manual User Manual

Page 53

Advertising
background image

53

LiMIT

(CtROL)

Code Display

Action

Choices

rL.

Use of relays:

• Limit function (LiMIT)

• Controller (CtROL)

Select using .

Press enter to proceed.

Note: Selecting CtROL leads to

Ct. controller menu group.

Lo

(Hi)

For limit 1 function principle,

see page 55. Select using .

Press enter to proceed.

Limit 1 contact response

N/C: normally closed contact

N/O: normally open contact

Select using .

Press enter to proceed

N/C

(N/O)

Limit 1 setpoint

Select using , edit number

using . Press enter to proceed.

00.00 pH

(

xx.xx pH)

Limit 1 hysteresis

Select using , edit number

using . Press enter to proceed.

00.50 pH

(

xx.xx pH)

Limit 1 delay

The contact is activated with

delay (deactivated without delay)

Select using , edit number

using . Press enter to proceed.

0010 sec

(

0 ... 9999

sec)

L1.