Hach-Lange SI794 P User Manual User Manual

Page 35

Advertising
background image

35

Code

4 - 20 mA

(0 - 20 mA)

Display

Action

Choices

o1.

Set output current range

Select using  key

Press enter to proceed

Current start

Enter lower end of scale,

depending on the process

variable selected (pH or ORP)

Select using , edit number

using . Press enter to proceed.

pH -2 ... 16

(-1500 mV

...

+1500mV)

Current end

Enter upper end of scale,

depending on the process

variable selected (pH or ORP)

Select using , edit number

using . Press enter to proceed.

pH -2 ... 16

(-1500 mV

...

+1500mV)

Output current

[pH]

20

4

14

0

Assignment of measured values:

Current start and current end

[pH]

20

4

7

5

[mA]

Example 1: Range pH 0 ... 14

Example 2: Range pH 5 ... 7 .

Advantage: Higher resolution in

range of interest

[mA]

Output current

Advertising