Stability check:

The measured mV value is

displayed.

To abort stability

check: Press cal.

(accuracy reduced)

Calibration with the first buffer

is terminated. Remove the

electrode and temp probe

from the first buffer solution

and rinse them thoroughly.

• One-point calibration:

Press cal to end the calibration.

Slope [%] and offset [mV] of

the electrode are displayed.

Press enter to proceed.

For one-point

calibration only:

• Two-point calibration:

Immerse sensor and tempera-

ture probe in the second buffer

solution.

Press enter to start.

The calibration pro-

cess runs again as

for the first buffer.

Retract electrode and temp

probe out of second buffer,

rinse off, re-install.

Repeat calibration: cal

End calibration: enter

Slope and offset of

electrode (related to

25 °C) are displayed.

pH value and “Hold” are

displayed alternately. Sensoface

is active, “enter” blinks.

Press enter to proceed.

Hold is deactivated after 20 sec.

Confirmation

prompt.