Passcodes – Hach-Lange SI794 P User Manual User Manual

Page 128

Advertising
background image

Key + Code

Description

Cal Info

Display of zero and slope

0000

1001

Product calibration

Adjusting the offset (zero point)

1105

Temp probe adjustment

1015

Configuration

Error Info

Display last error and erase

0000

Configuration
Sensor monitor

Display of uncorrected electrode potential (mV)

Parameter set 1/2

Selecting parameter set

1200

2222

7654

Key + Code

Description

Slope calibration

Adjusting the slope (sensor)

1100

5555

5556

5557

5559

Current source 1

Output current 1 specified

Relay test

Manual test of contacts

Manual controller

Manual specification of controller output

Current source 2

Output current 2 specified

Zero calibration (ISFET)

Default settings zero point (ISFET sensor only)

TA-194.130-HCE03 20081015

Software version: 1.x

Passcodes

(For fast access to the functions)

Calibration