Hach-Lange SI794 P User Manual User Manual
Page 75
Advertising
75
Converting slope [%] to slope [mV/pH] at 25 °C:
%
mV/pH
78
46.2
80
47.4
82
48.5
84
49.7
86
50.9
88
52.1
90
53.3
92
54.5
94
55.6
96
56.8
98
58.0
100
59.2
102
60.4
Converting offset to electrode zero point:
ZERO = 7 -
V
OFF
[mV]
S [mV / pH]
ZERO
V
OFF
S
Electrode zero point
Offset
Slope
Advertising