Automatic calibration with calimatic (buf -xx-) – Hach-Lange SI794 P User Manual User Manual

Page 70

Advertising
background image

70

Display

Action

Remark

Buffer recognition

While the “hourglass” icon is

blinking, the electrode and

temperature probe remain in

the first buffer solution.

Automatic calibration with Calimatic (BUF -xx-)

Automatic or manual temperature detection

Press cal, enter code: 1100

Select position using ,

edit number using ,

press enter to proceed.

Remove the electrode and

temperature probe, clean them,

and immerse them in the first

buffer solution (in any order).

When “Manual temp detection”

has been configured, enter the

value in the secondary display

using the arrow keys.

Press enter to start.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

The response time

of the electrode

and temperature

probe is consider-

ably reduced when

the electrode is

first moved about

in the buffer solu-

tion and then held

still.

Buffer recognition terminated,

the nominal buffer value is

displayed.

Transmitter in Hold

mode, measured

value frozen.

Sensoface inactive.

The transmitter can only operate properly when the buffer solutions used

correspond to the configured set. Other buffer solutions, even those with

the same nominal values, may demonstrate a different temperature

response. This leads to measurement errors.

Advertising