Error messages (error codes)

pH electrode

• Electrode defective

• Too little electrolyte in electrode

• Electrode not connected

• Break in electrode cable

• Incorrect electrode connected

• Measured pH value < -2 or > 16

• Measured ORP value

< -1999 mV or > 1999 mV

Error

Display

Problem

Possible causes

ERR 01

Measured

value

blinks

x

x

x

Redox electrode

• Electrode defective

• Electrode not connected

• Break in electrode cable

• Incorrect electrode connected

• Electrode potential < -1500 mV

• Electrode potential > 1500 mV

ERR 02

Measured

value

blinks

x

x

x

System error

Configuration or calibration data

defective; completely reconfigure

and recalibrate the transmitter.

Memory error in device program

ERR 98

“Conf”

blinks

x

x

x

x

Factory settings

EEPROM or RAM defective

This error message only occurs in

the case of a total defect.

The transmitter must be repaired

and recalibrated at the factory.

ERR 99

“FAIL”

blinks

x

x

x

x