Temp probe adjustment, Measurement – Hach-Lange SI794 P User Manual User Manual

Page 80

Advertising
background image

80

Display

Action

Remark

Enter measured temperature

value.

Select position usingkey,

edit number using key.

Press enter to proceed.

Press enter to exit adjustment.

HOLD will be deactivated after

20 sec.

Temp probe adjustment

Select calibration

(Press cal,

enter passcode: 1015)

Select position usingkey,

edit number using key.

Press enter to proceed.

Measure the temperature of

the process medium using an

external thermometer

Wrong settings

change the mea-

surement proper-

ties! If an invalid

code is entered, the

transmitter returns

to measuring mode.

Default:

Value of secondary

display.

The transmitter is in

the Hold mode.

Display

Remark

Measurement

In the measuring mode the main display shows the

configured process variable (pH or ORP [mV]) and the

lower display the temperature.

During calibration you can return to measuring mode

by pressing the cal key, during configuration by

pressing conf.

(Waiting time for signal stabilization approx. 20 sec).