Hach-Lange SI794 P User Manual User Manual

Page 37

Advertising
background image

37

0 sec

0...120 sec

Code Display

Action

Choices

o1.

Time interval of output filter

Default setting: 0 sec (inactive).

To specify a time interval:

Select position using ,

edit number using .

Press enter to proceed.

Time averaging filter

To smoothen the current output, a low-pass filter with adjustable

filter time can be switched on. When there is a jump at the input

(100 %), the output level is at 63 % after the time interval has been

reached.

The time interval can be set from 0 to 120 sec.

If the time interval is set to 0 sec, the current output follows the

input.

Note:

The filter only acts on the current output, not on the display, the

limit values, or the controller!

Time interval 0 to 120 sec

Advertising