o1.
Time interval of output filter
Default setting: 0 sec (inactive).
To specify a time interval:
Select position using ,
edit number using .
Press enter to proceed.
Time averaging filter
To smoothen the current output, a low-pass filter with adjustable
filter time can be switched on. When there is a jump at the input
(100 %), the output level is at 63 % after the time interval has been
reached.
The time interval can be set from 0 to 120 sec.
If the time interval is set to 0 sec, the current output follows the
input.
Note:
The filter only acts on the current output, not on the display, the
limit values, or the controller!
Time interval 0 to 120 sec