Specifications – Hach-Lange SI794 P User Manual User Manual

Page 98

Advertising
background image

98

Specifications

Output 2

0/4 ... 20 mA, max. 10 V, floating

(galv. connected to output 1)

Process variable

Temperature

Overrange*

22 mA in the case of temp error messages

Output averaging filter *

Low-pass, filter time interval 0 ... 120 s

Measurement error

1

< 0.3% current value + 0.05 mA

Start/end of scale *

20 ... 200 °C / -4 ... 392 °F

Span allowed

20 ... 220 K (36 ... 396 °F)

Power output

For operating an ISFET adapter

+3 V (V

0

= 2.9 ... 3.1 V / R

i

= 360 Ω

–3 V (V

0

= -4.8 ... –3,7 V / R

i

= 360 Ω

Alarm contact

Relay contact, floating

Contact ratings

AC< 250 V / < 3 A / < 750 VA

DC< 30 V / < 3 A / < 90 W

Contact response

N/C (fail-safe type)

Alarm delay

0000 ... 0600 sec

Limit values

Output via relay contacts R1, R2

Contacts R1, R2 floating but inter-connected

Contact ratings

AC< 250 V / < 3 A / < 750 VA

DC< 30 V / < 3 A / < 90 W

Contact response *

N/C or N/O

Delay *

0000 ... 9999 s

Setpoints *

As desired within range

Hysteresis *

00.00 ... 05.00 pH / 0000 ... 0500 mV

PID process controller

Output via relay contacts R1, R2 (see limit values)

(Relay R1: base valve, relay R2: acid valve)

Setpoint specification *

pH -02.00 ... 16.00 / -1500 ... +1500 mV

Neutral zone *

pH 00.00 ... 05.00 pH / 0000 ... 0500 mV

P-action component *

Controller gain KP: 0010 ... 9999 %

I-action component *

Reset time Tr:

0000 ... 9999 s

(0000 s = no integral action)

D-action component *

Rate time Td:

0000 ... 9999 s

(0000 s = no derivative action)

Controller type *

Pulse length or pulse frequency controller

Pulse period *

0001 ... 0600 s, min. ON time 0.5

(pulse length controller)