Hach-Lange SI794 P User Manual User Manual
Page 49

49
-01-BUF
(-02-BUF/
-03-BUF/
-04-BUF/
-05-BUF/
-06-BUF/
-07-BUF/
-08-BUF/
MAN/
DAT)
Code Display
Action
Choices
CA.
For pH measurement only:
Select calibration mode
BUF: Calibration with Calimatic
automatic buffer selection.
To do so, you must select your
buffer set:
-01- BUF: Mettler-Toledo
-02-BUF: Merck Titrisols, Riedel Fixanals
-03- BUF: Ciba (94)
-04-BUF: NIST technical buffers
-05-BUF: NIST standard buffers
-06-BUF: HACH buffers
-07-BUF: WTW technical buffers
-08- BUF: Hamilton Duracal
MAN: Calibration with
Manual Buffer Entry
DAT: Entry of offset and slope of
premeasured electrodes
Select using key, press enter to
proceed.
Enter calibration interval:
Entry of time interval within
which the transmitter is to be
calibrated.
With a time interval of 0000 hrs
the calibration timer is not active.
Select using , edit number
using. Press enter to proceed.
0000 h
(0000 ...
9999 h)