Overview of configuration steps, Menu code – Hach-Lange SI794 P User Manual User Manual

Page 30

Advertising
background image

30

Overview of configuration steps

PH/ORP

PH

(GLAS EL/FEt EL )

GLAS EL

0-20 mA/4-20 mA

4-20 mA

xxxx

00.00 pH

xxxx

14.00 pH

Time averaging filter

xxxx SEC

0 SEC

22 mA signal in the case of error

ON / OFF

OFF

Signal response during HOLD

Last / Fix

Last

Enter fixed value

xxx.x mA

021.0 mA

out2

Output 2

o2.

Select temperature unit

°C / °F

°C

Select temperature probe

100 Pt/1000 Pt/30NTC/

8.55NTC/3000B

1000 Pt

Select current range

0-20 mA/4-20 mA

4-20 mA

Enter current start

xxx.x

000.0 °C

Enter current end

xxx.x

0100.0 °C

Time averaging filter

0000 ... 0120 SEC)

0 SEC

22 mA signal for temp error

ON / OFF

OFF

Signal

response during HOLD

Last / Fix

Last

Enter fixed value

xxx.x mA

021.0 mA

tc.

Temperature compensation

tc.

Temp detection during meas

Auto/man (man: xxx.x °C)

Temp detection during cal

Auto/man (man: xxx.x °C)

Enter TC process medium

xx.xx %/K

00.00%/K

CAL

Calibration mode

CA.

Select calibration mode

BUF/MAN/DAT

06 BUF

Enter cal timer interval

xxxx h

0000 h

Menu

Code

out1

Output 1

o1.

Select process variable
Select electrode type
Select current range
Enter current start
Enter current end

Choices Default