Note: Characters represented in gray are blinking and can be edited.

Code Display

Action

Choices

o1.

Select configuration

(Press conf.)

Enter passcode “1200”

(Use  to select the position, use

to edit the number. When the

display reads “1200”, press enter

to confirm.)

The transmitter goes to HOLD

mode. (HOLD icon is active, red

LED blinks when “HOLD ON” has

been configured.)

Select parameter pH/ORP

Select using  key

Press enter to proceed

pH/ORP

Only with pH selected:

Select electrode type:

• Glass electrode

• ISFET electrode

Select using  key

Press enter to proceed

Glass

(FEt EL)

After correct input a

welcome text (CONF) is

displayed for approx. 3 sec