Code display action choices l2. hi, N/c (n/o), 00 ph – Hach-Lange SI794 P User Manual User Manual

Page 55: 50 ph, 0010 sec

Advertising
background image

55

Code Display

Action

Choices

L2.

Hi

(Lo)

For limit 2 function principle,

see Fig. below. Select using .

Press enter to proceed.

Limit 2 contact response

N/C: normally closed contact

N/O: normally open contact

Select using .

Press enter to proceed.

N/C

(N/O)

Limit 2 setpoint

Select using , edit number

using . Press enter to proceed.

14.00 pH

(

xx.xx pH)

Limit 2 hysteresis

Select using , edit number

using . Press enter to proceed.

00.50 pH

(

xx.xx %)

Limit 2 delay

The contact is activated with

delay (deactivated without delay)

Select using , edit number

using . Press enter to proceed.

0010 sec

(

0 ...

9999 sec)

Limit Lo

Hysteresis +

Setpoint

Contact

0

1

Signal

Limit Hi

Hysteresis -

Signal

Setpoint

Contact

0

1