Code display action choices l2. hi, N/c (n/o), 00 ph – Hach-Lange SI794 P User Manual User Manual
Page 55: 50 ph, 0010 sec
Advertising
55
Code Display
Action
Choices
L2.
Hi
(Lo)
For limit 2 function principle,
see Fig. below. Select using .
Press enter to proceed.
Limit 2 contact response
N/C: normally closed contact
N/O: normally open contact
Select using .
Press enter to proceed.
N/C
(N/O)
Limit 2 setpoint
Select using , edit number
using . Press enter to proceed.
14.00 pH
(
xx.xx pH)
Limit 2 hysteresis
Select using , edit number
using . Press enter to proceed.
00.50 pH
(
xx.xx %)
Limit 2 delay
The contact is activated with
delay (deactivated without delay)
Select using , edit number
using . Press enter to proceed.
0010 sec
(
0 ...
9999 sec)
Limit Lo
Hysteresis +
Setpoint
Contact
0
1
Signal
Limit Hi
Hysteresis -
Signal
Setpoint
Contact
0
1