Display

Action

Remark

Enter offset [mV].

Select position using ,

enter number using ,

press

enter to proceed.

Data entry of premeasured electrodes

Press cal, enter code: 1100

Select position using ,

edit number using ,

press enter to proceed.

Ready for calibration
Start with

enter.

If an invalid code is

entered, the trans-

mitter returns to

measuring mode.

Transmitter in Hold

mode, measured

value frozen.

Sensoface inactive.

You can directly enter the values for slope and offset of an electrode. The

values must be known, e.g. determined beforehand in the laboratory.

The DAT calibration mode must have been preset during configuration.

Enter slope [%]. Select using

, edit number using ,

press enter to proceed.

The transmitter displays the

new slope and offset (at 25 °C).

Press enter to proceed.

pH value and “Hold” are

displayed alternately. Sensoface

is active, “enter” blinks.

Press enter to proceed.

Hold is deactivated after 20 sec.

Confirmation

prompt.