Calibration – Hach-Lange SI794 P User Manual User Manual

Page 66

Advertising
background image

66

Calibration

Calibration adjusts the transmitter to the electrode.

Activation

Activate with cal

Enter passcode.

Use  and  to edit a

parameter.

Press enter to confirm/continue.

(To exit, press cal, then enter.)

HOLD

During cali-

bration the

transmitter

remains in

the Hold

mode.

During calibration the transmit-

ter remains in the Hold mode

for reasons of safety. Output

current is frozen (last value or

preset fixed value, depending

on configuration), limit and

alarm contacts are inactive. The

controller is in the configured

state, Sensoface is off, “Calibra-

tion” mode indicator is on. The

red LED blinks when “HOLD ON”

has been set.

HOLD icon

Input errors

The calibration parameters are

checked during the input. In the

case of an incorrect input “Err” is

displayed for approx. 3 sec. The

incorrect parameters cannot be

saved. Input must be repeated.

Exit

Press cal to exit. The measured

value is shown in the main

display alternately with “Hold”,

“enter” blinks, Sensoface is

active. Press enter key to end

the Hold mode. The measured

value is displayed. The output

current remains frozen for

another 20 sec (HOLD icon on,

“hourglass” blinks).