Zero adjustment (isfet) – Hach-Lange SI794 P User Manual User Manual

Page 68

Advertising
background image

68

Display

Action

Remark

If the zero offset of

the electrode is too

large (> ± 200 mV),

the CAL ERR error

message is generat-

ed. In that case the

electrode cannot be

calibrated.

Zero adjustment (ISFET)

Allows use of electrodes with differing nominal zero (pH only)

Press cal, enter code: 1001

Select position using ,

edit number using ,

press enter to proceed.

Immerse electrode in a

pH 7.00 buffer. Enter the tem-

perature-corrected pH value in

the range 6.50 to 7.50 using the

arrow keys (see buffer table).

Press enter to confirm.

The transmitter is in

the Hold mode.

If an invalid code is

entered, the

transmitter returns

to measuring mode.

Ready for calibration

“CAL” and “enter” are blinking.

Display (3 sec)

Stability check.

The measured value [mV] is

displayed.

The “hourglass” icon is blinking.

Note:

Stability check can

be stopped (by

pressing cal).

However, this

reduces calibration

accuracy.

Advertising