Product calibration – Hach-Lange SI794 P User Manual User Manual

Page 76

Advertising
background image

76

Product calibration

Calibration by sampling

During product calibration the electrode remains in the process.

The measurement process is only interrupted briefly.

Procedure: During sampling the currently measured value is saved

in the transmitter. The transmitter immediately returns to measuring

mode.

The calibration mode indicator blinks and reminds you that calibra-

tion has not been terminated.

The sample is measured in the lab or directly on the site using a

portable meter. To ensure an exact calibration, the sample tempera-

ture should correspond to the measured process temperature.

The measured sample value is then entered in the transmitter.

From the difference between the saved measured value and entered

sample value, the transmitter calculates the new offset (one-point

calibration).

If the sample is invalid, you can take over the value saved during

sampling. In that case the old calibration values are saved.

Afterwards, you can start a new product calibration.

Display

Action

Remark

Product calibration step 1:

Press cal, enter code: 1105

(Presskey to select position,

enter number usingkey,

press enter to confirm)

Take sample and save value.

Press enter to proceed.

If an invalid code

is entered, the

transmitter returns

to measuring mode.

Now the sample

can be measured in

the lab.