60

Selecting parameter set 1/2

Manually or via a signal at the Control input

Display

Action

Choices

Select:

• Parameter set 1 (MAN)

• Parameter set 2 (MAN)

• Automatic switchover via

control input (Ctr-EXT)

Select using , press enter to proceed.

To select a parameter set, press conf,

enter code: 7654

Select position using , edit number

using , press enter to proceed.

Wrong settings change the measurement

properties! If an invalid code is entered, the

transmitter returns to measuring mode.

With -1- or -2- selected:

Since the complete device configura-

tion is changed in one step, there is a

security prompt (No/Yes).

Note:

When pressing enter directly,

the selection is not stored.

Activation of parameter set 2 is

indicated by “88” in the upper left

corner of the display.

After correct input a

welcome text (CONF) is

displayed for approx. 3 sec

-1-

(-1- MAN

-2-MAN

Ctr-EXT)

For Ctr-EXT,

see next

page