Selecting parameter set 1/2 – Hach-Lange SI794 P User Manual User Manual
Page 60

60
Selecting parameter set 1/2
Manually or via a signal at the Control input
Display
Action
Choices
Select:
• Parameter set 1 (MAN)
• Parameter set 2 (MAN)
• Automatic switchover via
control input (Ctr-EXT)
Select using , press enter to proceed.
To select a parameter set, press conf,
enter code: 7654
Select position using , edit number
using , press enter to proceed.
Wrong settings change the measurement
properties! If an invalid code is entered, the
transmitter returns to measuring mode.
With -1- or -2- selected:
Since the complete device configura-
tion is changed in one step, there is a
security prompt (No/Yes).
Note:
When pressing enter directly,
the selection is not stored.
Activation of parameter set 2 is
indicated by “88” in the upper left
corner of the display.
After correct input a
welcome text (CONF) is
displayed for approx. 3 sec
-1-
(-1- MAN
-2-MAN
Ctr-EXT)
For Ctr-EXT,
see next
page