Beeper gasket – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 118

Advertising
background image

E Series Service Manual

5-8

9650-1450-01 Rev. D

10. Cut video cable tie wrap.

11. Push on board retaining clip to release the control board

from front panel frame, but do not remove board until next
step.

Note:

When reinstalling see Table 5.1 at end of section for setting
control board dip switches.

12. As you lift the control board, beeper gasket may need to be cut

from the front panel membrane. Use an exacto blade.

Beeper Gasket

Note:

If unit has audio option, remove microphone from panel and
disconnect from board

Note:

Below is a picture of the opposite side of control board
showing the beeper gasket.