ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 59

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

2-7

CONFIRM MANUAL MODE

Displayed when manual mode is entered. Alerts user to
confirm that manual mode is desired.

DEFIB DISABLED

User prompt issued simultaneously with other faults if
defib is disabled.

Possible configuration problem. Replace high
voltage module. Call ZOLL Technical Support.

DEFIB FAULT 71

More than 50 internal dumps occurred in less than 20
minutes.

Turn the unit to OFF and back on. If fault persists,
replace high voltage module.

DEFIB FAULT 72

Charging inhibited, voltage out of range.

Turn the unit to OFF and back on. If fault persists,
replace high voltage module.

DEFIB FAULT 76

Capacitor voltage too high for selected energy.

Replace high voltage module or capacitor.

DEFIB FAULT 77

Capacitor voltage > than absolute rated max.

Replace high voltage module or capacitor.

DEFIB FAULT 78

Unable to charge defib cap.

Replace high voltage module or capacitor.

DEFIB FAULT 79

Defibrillator charging too slowly.

Replace high voltage module or capacitor.

DEFIB FAULT 80

4 defibrillator faults detected within 20 second period.

Replace high voltage module or capacitor.

DEFIB FAULT 81

Discharge switch in undefined state.

Replace high voltage module or capacitor.

DEFIB FAULT 84

“Upper” discharge transistor shorted (measured via
applicable A/D channel).

Replace high voltage module.

DEFIB FAULT 85

“Lower” discharge transistor shorted (measured via
applicable A/D channel).

Replace high voltage module.

DEFIB FAULT 86

One discharge switch closed during power up test.

Replace paddles, control board or system board.

DEFIB FAULT 87

Both discharge switches closed during power up test.

Replace paddles, control board or system board.

DEFIB FAULT 94

Processor fault causing safety monitor port to be non-
functional.

Replace system board, high voltage module or
capacitor.

DEFIB FAULT 95

Safe or shutdown line is not functional.

Replace high voltage module.

DEFIB FAULT 108

VMON voltage is less than the target energy during
charging.

Replace high voltage module or capacitor.

Error Message

Explanation

User
Advisory

Technical Action