Warning – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 228

Advertising
background image

E Series Operator’s Guide

11-2

9650-1210-01

Rev.

R

Testing the Defibrillator (Semiautomatic Mode)

Perform these tests on all AED units periodically.

Power-Up Sequence Test

Follow these steps to verify the unit’s power-up
sequence.

1. Turn the selector switch on the front panel to the ON

position

2. Verify the following:

The unit emits a 4-beep tone to indicate the
power-up self test has been successfully
completed.

The CHECK PADS message is displayed and
announced.

Defibrillator Test

Perform the defibrillator test to verify that the unit
analyzes ECG rhythm and delivers shock treatment
correctly. To perform this test, you need an ECG rhythm
simulator.

1. Ensure that the selector switch on the E Series front

panel is in the OFF position.

2. Connect the patient end of the multifunction cable to

the defibrillator test connector.

3. Turn the selector switch to ON.

4. Press the ANALYZE button and verify that the unit

charges to 30 J (30 J READY message).

5. Once unit has charged, verify the SHOCK button

illuminates.

6. Press and hold the SHOCK button.

7. Verify that the unit briefly displays the TEST OK

message and prints a stripchart.

This message indicates that the unit delivered energy
within specifications.

Note: If TEST FAILED message appears, contact the

ZOLL Technical Service Department immediately.

8. Connect the multifunction cable to the ZOLL ECG

Simulator, and set the simulator to VF.

9. Verify that within 30 seconds the unit displays and

and announces the CHECK PATIENT prompt.

10. Press the ANALYZE button on the front panel and

verify the unit charges to 120 J or other preconfigured
level.

11. Once unit has charged, verify that it illuminates the

SHOCK button and that it displays and announces
the PRESS SHOCK prompt.

12. Press and hold the SHOCK button and verify unit

discharges.

Note: If a LOW BATTERY message appears during

testing at the beginning of a shift, the battery
currently in use is close to depletion; replace and
charge the battery. The device does not test the
battery for adequate charge to support extended
use of the unit; capacity can only be determined
by testing the battery in a Base Power
Charger 4x4 or SurePower Charger.

Testing the Defibrillator (Manual Mode)

Perform these tests on all standard E Series units
periodically.

Note: During the Energy Delivery Test, the unit

discharges only when the energy level is set to
30 joules.

Power-Up Sequence Test

Follow these steps to verify the unit’s power-up
sequence. To perform this test, you need an ECG rhythm
simulator.

1. Ensure the selector switch on the E Series front

panel is set to OFF, and connect the ECG monitoring
cable to the simulator.

2. Turn the selector switch to the MONITOR position.
3. Verify that the:

The unit emits a 4-beep tone to indicate the
power-up self test has been successfully
completed.

ECG size is x1

MONITOR message displays on the LCD

ECG source is PADDLES or PADS

If no ECG cable is connected to the simulator, the
message ECG LEAD OFF is displayed and the ECG
display shows a dashed line instead of a solid line.

Delivered Energy and Shock Buttons

Perform this test at the start of each shift.

Paddles Setup:

1. Verify adult paddles are installed and are inserted all

the way into their holders on the side of the E Series
unit.

2. Turn the selector switch to DEFIB.
3. Set the defibrillator energy level to 30 joules.
4. Press the CHARGE button on the apex handle and

verify that the unit displays the DEFIB 30J READY
message.

WARNING

When performing this test using paddles, use your
thumbs to operate the SHOCK buttons in order to
avoid an inadvertent shock. No portion of the hand
should be near the paddle plates.