Isolated Power Supply Removal
1. Disconnect the molex connector.
2. Disconnect the 2 spade connectors.
3. Remove the 4 screws securing the board.
4a (Old Style MS-11) Carefully lift the isolated power supply
and remove long grey stranded cable (EtCO2), and the 2
flat cables, then remove board.
Note:
During reassembly route grey cable as shown in picture.