Isolated Power Supply Removal

1. Disconnect the molex connector.

2. Disconnect the 2 spade connectors.

3. Remove the 4 screws securing the board.

4a (Old Style MS-11) Carefully lift the isolated power supply

and remove long grey stranded cable (EtCO2), and the 2
flat cables, then remove board.

Note:

During reassembly route grey cable as shown in picture.