0 co, Accuracy check (optional test to check etco, Sensor) – ZOLL E Series Monitor Defibrillator Rev D User Manual

Page 43

Advertising
background image

E Series Service Manual

9650-1450-01 Rev. D

1-33

20.0 CO

2

Accuracy Check (Optional test to check EtCO

2

sensor)

Tools Needed: CAPNOSTAT 5 Mainstream cable with airway adapter (or CAPNOSTAT 5 Sidestream cable with cannula), Gas regulator, 5% calibration
gas, Thermometer.

Test Setup: None

Do this...

Observe this...

Pass/Fail/NA

20.1

Connect the CAPNOSTAT 5 CO2 Mainstream
cable with airway adapter attached, or the
CAPNOSTAT 5 Sidestream cable with the
cannula attached, to the yellow connector at the
back of the E Series.

20.2

While pressing and holding the second softkey
from the left, turn the selector switch to Monitor
(ON for AED units).

The unit displays EtCO

2

Calibration screen.

20.3

Wait for the sensor to warm up.

The message WARM UP is displayed for approximately one minute.

20.4

Obtain current room temperature in

Centigrade (Cº).

20.5

Press the Select Gas Temp softkey to enter the
CO

2

Accuracy screen

20.6

Use the Prev, Next, Inc and Dec softkeys to set
each digit of the gas temperature parameter in
the CAPNOSTAT 5 CO

2

Sensor until Gas

Degrees C is equal to the room temperature.

20.7

Press the Return softkey to store the
temperature and return to the main EtCO

2

Calibration screen.

o o o

20.8

If using a mainstream CAPNOSTAT 5 CO

2

Sensor/Airway Adapter, press the Zero softkey
to zero the sensor.

Note: Sidestream sensors do not require
zeroing, they auto-zero.

o o o